HRC Flight Builds


BUILD #1 BUILD #2

HRC FLIGHT BUILD #2


IMAGES GRAPHS

CHIP NAME: 7282LDN01-02
CTE  (Fe55)
serial / parallel
READ NOISE
*
e-rms
 
A
0.9999947 / 0.9999818
   
A
5.4
 
B
0.9999901 / 0.9999880
   
B
5.2
 
C
0.9999951 / 0.9999853
   
C
5.1
 
D
0.9999938 / 0.9999807
   
D
5.4
DARK CURRENT
 
20 e-/pix/hr
 
FULL WELL
 
150,000

* The readnoise reported is limited by test equipment

l (nm)
OTA
M1-2
M3
window
loss
factor
HRC #2
QE
HRC #2
NET QE
250
0.60
0.762
0.923
0.956
0.95
0.75
0.29
300
0.61
0.782
0.906
0.964
0.95
0.73
0.29
350
0.62
0.802
0.886
0.961
0.95
0.61
0.24
400
0.64
0.810
0.869
0.960
0.95
0.61
0.25
500
0.64
0.810
0.857
0.955
0.95
0.61
0.25
600
0.64
0.801
0.856
0.953
0.95
0.65
0.26
700
0.62
0.778
0.849
0.951
0.95
0.64
0.24
800
0.60
0.723
0.820
0.935
0.95
0.50
0.16
900
0.64
0.780
0.861
0.936
0.95
0.33
0.13
1000
0.72
0.870
0.924
0.937
0.95
0.14
0.07

top


HRC FLIGHT BUILD #1


IMAGES GRAPHS

CHIP NAME: 7161LXD01-01
CTE  (Fe55)
serial / parallel
READ NOISE
 *
e-rms
 
A
0.9999961 / 0.9999885
   
A
4.5
 
B
0.9999987 / 0.9999811
   
B
4.6
 
C
0.9999966 / 0.9999843
   
C
4.6
 
D
0.9999974 / 0.9999821
   
D
4.6
DARK CURRENT
 
20 e-/pix/hr
 
FULL WELL
 

* The readnoise reported is limited by test equipment

l (nm)
OTA
M1-2
M3
window
loss
factor
HRC #2
QE
HRC #2
NET QE
250
0.60
0.762
0.923
0.956
0.95
0.35
0.13
300
0.61
0.782
0.906
0.964
0.95
0.36
0.14
350
0.62
0.802
0.886
0.961
0.95
0.39
0.16
400
0.64
0.810
0.869
0.960
0.95
0.47
0.19
500
0.64
0.810
0.857
0.955
0.95
0.59
0.24
600
0.64
0.801
0.856
0.953
0.95
0.63
0.25
700
0.62
0.778
0.849
0.951
0.95
0.62
0.23
800
0.60
0.723
0.820
0.935
0.95
0.50
0.16
900
0.64
0.780
0.861
0.936
0.95
0.35
0.13
1000
0.72
0.870
0.924
0.937
0.95
0.14
0.07